Soylak Mustafa, Divrikli Umit, Saracoglu Sibel, Elci Latif
Erciyes Universitesi, Fen-Edebiyat Fakultesi, Kimya Bolumu, 38039, Kayseri, Turkey.
Environ Monit Assess. 2007 Apr;127(1-3):169-76. doi: 10.1007/s10661-006-9271-0. Epub 2006 Oct 21.
Cellulose nitrate membrane filter was used for the preconcentration-separation of Cu, Co, Cd, Pb and Cr ions. The analyte ions were collected on the membrane filter by the aid of carmine. Then membrane filter was dissolved by using nitric acid. The levels of the analytes in the final solutions were determined by flame atomic absorption spectrometry (FAAS). The analytical parameters including pH, amounts of carmine, sample volumes etc. have been optimized. No influences have been observed from the matrix ions. The detection limits for analytes were in the range of 0.08 microg/l-0.93 microg/l. The validation of the procedure was checked by the analysis of standard reference sediment (GBW 07309). The present method has been successfully applied for the FAAS determinations of analyte ions in real samples including black tea and magnesium salts.
硝酸纤维素膜滤器用于铜、钴、镉、铅和铬离子的预富集分离。借助洋红将分析物离子收集在膜滤器上。然后用硝酸溶解膜滤器。通过火焰原子吸收光谱法(FAAS)测定最终溶液中分析物的含量。对包括pH值、洋红用量、样品体积等分析参数进行了优化。未观察到基体离子的影响。分析物的检测限在0.08微克/升 - 0.93微克/升范围内。通过分析标准参考沉积物(GBW 07309)对该方法进行了验证。本方法已成功应用于红茶和镁盐等实际样品中分析物离子的FAAS测定。