Liu Rui, Liang Pei
Key Laboratory of Pesticide & Chemical Biology of Ministry of Education, College of Chemistry, Central China Normal University, Wuhan 430079, PR China.
Anal Chim Acta. 2007 Dec 5;604(2):114-8. doi: 10.1016/j.aca.2007.10.008. Epub 2007 Oct 11.
A new method has been developed for the determination of gold based on separation and preconcentration with a microcolumn packed with nanometer TiO2 immobilized on silica gel (immobilized nanometer TiO2) prior to its determination by flame atomic absorption spectrometry. The optimum experimental parameters for preconcentration of gold, such as pH of the sample, sample flow rate and volume, eluent and interfering ions, have been investigated. Gold could be quantitatively retained by immobilized nanometer TiO2 in the pH range of 8-10, then eluted completely with 0.1 mol L(-1) HNO3. The detection limit of this method for Au was 0.21 ng mL(-1) with an enrichment factor of 50, and the relative standard deviation (R.S.D.) was 1.8% at the 100 ng mL(-1) Au level. The method has been applied for the determination of trace amounts of Au in geological and water samples with satisfactory results.
已开发出一种测定金的新方法,该方法基于在通过火焰原子吸收光谱法测定金之前,用填充有固定在硅胶上的纳米二氧化钛(固定化纳米二氧化钛)的微柱进行分离和预富集。研究了金预富集的最佳实验参数,如样品的pH值、样品流速和体积、洗脱液和干扰离子。固定化纳米二氧化钛能在pH值为8 - 10的范围内定量保留金,然后用0.1 mol L(-1)硝酸将其完全洗脱。该方法对金的检测限为0.21 ng mL(-1),富集倍数为50,在100 ng mL(-1)金水平下相对标准偏差(R.S.D.)为1.8%。该方法已应用于地质和水样中痕量金的测定,结果令人满意。