The CNAP HD Monitor (CNSystems, Graz, Austria) and the ccNexfin (The ClearSight System: Edwards Lifesciences Corporation, Irvine, CA) are continuous, noninvasive blood pressure monitors using a finger-application device. These devices show a promising ability to allow for rapid detection of hemodynamic derangement when compared with oscillometry. The accuracy and precision of these devices as blood pressure monitors has been evaluated when compared with intra-arterial catheters. Additionally, they can be used to measure beat-to-beat cardiac output (CO). As CO monitors, they are capable of trending changes in CO when compared with a transpulmonary thermodilution monitor. Difficulty with use in critically ill and awake patients has been encountered because of altered microvascular physiology and patient movement. The principles of operation and clinical validation of these devices are presented. The clinicians who are interested in using these devices in their clinical setting should be aware of the relatively large bias and CIs in the hemodynamic measurements.
CNAP HD监护仪(奥地利格拉茨的CNSystems公司)和ccNexfin(ClearSight系统:加利福尼亚州尔湾市的爱德华兹生命科学公司)是使用手指应用装置的连续无创血压监测仪。与示波法相比,这些装置在快速检测血流动力学紊乱方面显示出有前景的能力。与动脉内导管相比时,已评估了这些装置作为血压监测仪的准确性和精密度。此外,它们可用于测量逐搏心输出量(CO)。作为CO监测仪,与经肺热稀释监测仪相比时,它们能够追踪CO的变化。由于微血管生理改变和患者活动,在危重症和清醒患者中使用时遇到了困难。介绍了这些装置的操作原理和临床验证情况。有兴趣在临床环境中使用这些装置的临床医生应了解血流动力学测量中相对较大的偏差和可信区间。