Suppr超能文献

新生儿高血压

Neonatal Hypertension

作者信息

Pattnaik Priyam, Shah Manan, Verma Rita P.

机构信息

University of Connecticut School of Medicine & Connecticut Children's

Rutgers University

Abstract

Despite its presence in clinical practice for the past 4 to 5 decades, neonatal hypertension has only recently gained recognition as a distinct neonatal morbidity. The absence of comprehensive normative data on neonatal blood pressure hindered earlier efforts to evaluate and manage this condition. Limitations in the availability of accurate invasive and noninvasive measurement techniques further complicated assessment. Additionally, significant variability in blood pressure based on gestational age, postnatal age, birth weight, and gender within the neonatal population created challenges in defining normal ranges and establishing a standardized definition of neonatal hypertension. The clinical trajectory, long-term outcomes, and potential sequelae of neonatal hypertension remain incompletely understood. No clear consensus has emerged regarding the appropriate use or selection of antihypertensive medications in affected infants. Current knowledge continues to evolve, with ongoing efforts to clarify the definition, risk factors, etiopathogenesis, and management strategies for hypertension in newborns. Diagnosis of neonatal hypertension requires systolic or diastolic blood pressure values at or above the 95th percentile for postconceptual age, recorded on three separate occasions. Blood pressure values exceeding the 99th percentile suggest severe hypertension and warrant the initiation of antihypertensive treatment, along with further investigation to determine the underlying cause (see . Neonatal Normative Blood Pressure Data).  Blood pressure can be measured via invasive or noninvasive methods. The invasive intra-arterial blood pressure measurement and continuous monitoring, regarded as the gold standard, are performed by utilizing an indwelling catheter in the umbilical, radial, or posterior tibial arteries, which is connected to a pressure transducer and, via that, to a multichannel display patient monitor. This method is generally reserved for sick, unstable, or extremely premature infants. The following steps and precautions should be taken while measuring the blood pressure intra-arterially via the transducer: The transducer must remain positioned at the level of the heart to reflect actual pressure values. Air bubbles in the tubing must be avoided, as their presence can elevate diastolic pressure and lower systolic pressure, distorting the readings. A clear dicrotic notch should appear on the arterial waveform to confirm correct waveform representation. Tubing should consist of low-compliance material and maintain the shortest acceptable length, since longer tubing can falsely reduce measured values. The pressure transducer must be referenced to zero at atmospheric pressure to calibrate correctly. A continuous heparin infusion should be used to irrigate the transducer and maintain line patency, thereby reducing the risk of clot formation. The umbilical catheter must be appropriately sized, as narrow catheters may underestimate systolic pressure. Removal of the umbilical artery catheter is recommended after 5 to 7 days of use. Prolonged catheterization increases the risk of thrombus formation, which may contribute to inaccurate readings and potential complications.   Automated oscillometry is the most common and widely used noninvasive method for measuring blood pressure in the neonatal intensive care unit. This device detects the maximum blood pressure oscillations from arterial blood flow as mean blood pressure, which is then converted into projected systolic and diastolic blood pressures using standard proprietary algorithms. Oscillometric blood pressure measurements generally correlate well with invasive intra-arterial readings. However, automated oscillometry may overestimate intra-arterial systolic blood pressure values by 3 to 8 mm Hg, thereby overdiagnosing hypertension, and becomes inaccurate when mean arterial pressure drops below 30 mm Hg, thus missing hypotension. The device may also underestimate systolic blood pressure in small-for-gestational-age infants. Variations in blood pressure accuracy across studies might stem from the fact that each manufacturer of oscillometric devices employs a unique algorithm to calculate blood pressure, and several studies have highlighted discrepancies among different devices. For accuracy, the optimum cuff width is suggested to be in a ratio of 0.44 to 0.55, with the arm circumference covering approximately 80% of the patient's arm length. The size should be standardized for uniformity in the results. The blood pressure becomes erroneously high if the cuff size is too small. At the time of measurement, the infant should lie supine, quietly awake, calm, preferably sleeping, and about 1 to 1.5 hours postprandial. At least three readings, 2 minutes apart, should be taken in the right arm as the preferred site. Using a sphygmomanometer is not recommended because the Korotkoff sounds are not loud enough to be reliably audible in this age group of infants. Ultrasound Doppler is rarely used as a regular blood pressure monitoring device, as it can underestimate systolic blood pressure values.

摘要

尽管在过去4至5十年的临床实践中就已遇到新生儿高血压,但直到最近它才被作为一种独特的新生儿疾病而得到认识和研究。由于缺乏准确的有创或无创测量技术,早期疾病过程的评估和管理因缺乏新生儿血压(BP)的全面规范数据而受到影响。在转诊人群中,血压随胎龄、出生后年龄、出生体重和性别存在显著且快速的变异性,这给确定正常值和新生儿高血压(NH)的包容性定义带来了进一步挑战。NH的临床病程、结局和长期后遗症仍未完全明了,专家们在用于治疗的降压药物的使用和选择上也缺乏共识。 本主题回顾了有关新生儿高血压的定义、危险因素、病因发病机制和管理的当前信息。当在3个不同时间测量的收缩压或舒张压(DBP)值超过或等于婴儿孕龄的第95百分位数时,即可诊断为NH。收缩压(SBP)超过第99百分位数提示严重高血压,表明需要开始抗高血压治疗并进行特定检查以确定病因发病机制。新生儿血压的规范数据列于表中(见新生儿人群出生后2周时的血压)。 (改编自参考文献11,迪翁等人,《儿科肾病学》,2012年;27:159 - 60)血压可通过有创或无创方法测量。有创动脉内血压测量和连续监测被视为金标准,通过在脐动脉、桡动脉或胫后动脉之一中留置导管来完成,该导管连接到压力传感器,再通过压力传感器连接到多通道显示患者监护仪。这种方法通常仅用于病情严重和不稳定的婴儿以及极早产儿。通过传感器进行动脉内血压测量时应采取以下步骤和注意事项。传感器应置于心脏水平。管道中不应有气泡,因为气泡的存在可能会使DBP升高并使SBP降低。动脉波形上应可见重搏波切迹。管道应具有低顺应性且长度应为可接受的最小长度,因为增加长度会错误地降低测量值。压力传感器应以零大气压为参考点。应通过持续输注肝素对传感器进行冲洗。脐导管应尺寸合适。导管过窄会错误地降低SBP。脐动脉导管应在5至7天后取出,因为长时间置管可能会增加血栓形成的风险,导致读数错误。 自动振荡法是新生儿重症监护病房中最常用的无创血压测量方法。该设备通过检测动脉血流中的最大血压振荡来获取平均血压(MBP),然后使用标准专有算法将其转换为预计的收缩压和舒张压。振荡法测量的血压通常与有创动脉内测量值相关性良好。然而,它可能会使动脉内SBP值高估3至8毫米汞柱,从而过度诊断高血压,并且当平均动脉压(MAP)降至30毫米汞柱以下时会变得不准确,进而漏诊低血压。该设备在小于胎龄儿中也可能低估SBP。为确保准确性,建议最佳袖带宽度与手臂周长的比例为0.45至0.70,手臂周长应覆盖手臂长度的80%,并且尺寸应标准化以确保结果的一致性。如果袖带尺寸过小,血压会错误地偏高。 在测量时,婴儿应仰卧,安静清醒、平静,最好处于睡眠状态,且在餐后约1至1.5小时。应首选右臂,至少间隔2分钟测量3次读数。不建议使用血压计,因为在这个年龄组的婴儿中柯氏音不够响亮,无法可靠地听到。超声多普勒很少用作常规血压监测设备,因为它可能会低估SBP值。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验