Magni Nico, Olds Margie, McLaine Sally
Department of Physiotherapy, School of Clinical Sciences, Auckland University of Technology, Auckland, New Zealand.
Auckland Shoulder Clinic, Auckland, New Zealand.
Hand Ther. 2023 Mar;28(1):33-39. doi: 10.1177/17589983231152958. Epub 2023 Feb 20.
Digital dynamometers to assess grip strength are becoming more common in research and clinical settings. The aim of the study was to assess validity and reliability of the K-force dynamometer compared to the Jamar dynamometer. We also aimed to assess differences over the course of three measurements.
Twenty-seven healthy participants were included. Three trials with the K-force and Jamar dynamometers were completed. Testing order was randomised. Intraclass correlation coefficients (ICCs) with absolute agreement assessed reliability and validity. Standard error of the measurement (SEM) and minimal detectable change (MDC) were calculated. Concurrent validity was assessed using Pearson's correlations and ICCs. Differences between the three repetitions were assessed using one-way repeated measures ANOVAs.
Both the K-force and the Jamar presented excellent intra-rater reliability with ICCs ranging from 0.96 to 0.97. The SEM ranged from 1.7 to 2 kg and the MDC from 4.7 to 5.7 kg for both dynamometers. The concurrent validity of the K-force was high ( ≥ 0.89). However, the K-force underestimated the grip strength by 4.5-8.5 kg. There was no change in grip strength with either dynamometer over the course of three trials.
The K-force is reliable, but it underestimates grip strength by 4.5-8.5 kg compared to the Jamar dynamometer. K-force can be used to monitor progress over time but cannot be used to compare results against normative data. The use of a single measurement when assessing grip strength is sufficient when assessing healthy subjects.
在研究和临床环境中,用于评估握力的数字测力计越来越普遍。本研究的目的是评估K型测力计与Jamar测力计相比的有效性和可靠性。我们还旨在评估三次测量过程中的差异。
纳入27名健康参与者。使用K型和Jamar测力计完成三次试验。测试顺序是随机的。使用具有绝对一致性的组内相关系数(ICC)评估可靠性和有效性。计算测量标准误差(SEM)和最小可检测变化(MDC)。使用Pearson相关性和ICC评估同时效度。使用单向重复测量方差分析评估三次重复之间的差异。
K型和Jamar测力计均表现出出色的评分者内可靠性,ICC范围为0.96至0.97。两种测力计的SEM范围为1.7至2千克,MDC范围为4.7至5.7千克。K型测力计的同时效度较高(≥0.89)。然而,K型测力计低估握力4.5至8.5千克。在三次试验过程中,两种测力计的握力均无变化。
K型测力计可靠,但与Jamar测力计相比,它低估握力4.5至8.5千克。K型测力计可用于监测随时间的进展,但不能用于将结果与标准数据进行比较。在评估健康受试者时,评估握力时使用单次测量就足够了。