Suppr超能文献

一项评估新型LubriShield™ 福来导尿管安全性的前瞻性随机试点研究:一种带有永久性涂层的留置导尿管。

A prospective randomised pilot study evaluating the safety of the novel LubriShield™ Foley catheter: a permanently coated indwelling urinary catheter.

作者信息

Andreasson Anders, Andersson Jan, Larsson Henrik, Ekerhult Teresa, Stranne Johan

机构信息

Department of Urology, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden; Department of Urology, Region Västra Götaland, Sahlgrenska University Hospital, Gothenburg, Sweden.

Department of Medicine, Karolinska Institutet, Stockholm, Sweden.

出版信息

Scand J Urol. 2025 Jun 24;60:124-129. doi: 10.2340/sju.v60.43994.

Abstract

OBJECTIVE

Catheter-associated urinary tract infections (CAUTIs) are prevalent healthcare-associated infections, arising from biofilm-forming bacteria. This may be prevented by coating the catheter with an antifouling substance. The novel LubriShield™ Foley catheter is coated with a superhydrophilic surface and a covalently bonded antifouling ligand. Preclinical studies revealed that the coating established a persistent local antifouling environment, inhibiting uropathogenic bacteria from forming biofilms. No substance release has been detected from the coating. The coating achieved a 28-fold reduction in surface friction compared to an uncoated catheter. The aim of this study is to assess the clinical safety of the catheter in patients.

MATERIALS & METHODS: In a prospective single-centre randomised study, 30 patients undergoing transurethral resection of bladder tumour were enrolled and randomly assigned to receive either a standard control catheter or the novel LubriShield™ catheter. Urinary cultures were obtained twice. The duration of catheterisation for the patients ranged from 3 to 24 h. The primary outcome was the assessment of device-specific adverse events (AEs). Secondary outcomes included evaluations of pain, irritation and discomfort, measured using the Numeric Rating Scale (NRS) (0-10) via a patient questionnaire.

RESULTS

There were no serious adverse events (SAEs) or AEs reported for the coated catheters. Urinary cultures showed no significant differences between the coated and uncoated catheters. Both patients and healthcare professionals rated the NRS equally for the two types of catheters.

CONCLUSIONS

The novel-coated LubriShield™ catheter was found to be safe for short-term clinical use.

摘要

目的

导尿管相关尿路感染(CAUTIs)是常见的医疗相关感染,由形成生物膜的细菌引起。这可以通过用防污物质涂覆导尿管来预防。新型LubriShield™Foley导尿管涂有超亲水表面和共价结合的防污配体。临床前研究表明,该涂层建立了持久的局部防污环境,抑制尿路致病性细菌形成生物膜。未检测到涂层有物质释放。与未涂覆的导尿管相比,该涂层使表面摩擦力降低了28倍。本研究的目的是评估该导尿管在患者中的临床安全性。

材料与方法

在一项前瞻性单中心随机研究中,纳入30例接受膀胱肿瘤经尿道切除术的患者,并随机分配接受标准对照导尿管或新型LubriShield™导尿管。进行了两次尿培养。患者的导尿持续时间为3至24小时。主要结局是评估特定器械不良事件(AE)。次要结局包括通过患者问卷使用数字评分量表(NRS)(0 - 10)对疼痛、刺激和不适进行评估。

结果

对于涂覆导尿管,未报告严重不良事件(SAE)或AE。尿培养显示涂覆导尿管和未涂覆导尿管之间无显著差异。患者和医护人员对两种类型导尿管的NRS评分相同。

结论

新型涂覆的LubriShield™导尿管被发现短期临床使用安全。