使用焦点移动型可调节人工晶状体客观测量人工晶状体的运动和屈光变化。

Objective measurement of intraocular lens movement and dioptric change with a focus shift accommodating intraocular lens.

作者信息

Hancox Joanne, Spalton David, Heatley Catherine, Jayaram Hari, Marshall John

机构信息

Ophthalmology Department, St. Thomas' Hospital, London, United Kingdom.

出版信息

J Cataract Refract Surg. 2006 Jul;32(7):1098-103. doi: 10.1016/j.jcrs.2006.01.092.

Abstract

PURPOSE

To objectively measure the shift and refractive change of the 1CU accommodating intraocular lens (IOL) (HumanOptics) and compare them to that of a monofocal AcrySof MA30 IOL (Alcon Laboratories) in the fellow eye.

SETTING

Ophthalmology Department, St. Thomas' Hospital, London, United Kingdom.

METHODS

Thirty patients who had bilateral cataract surgery 18 to 24 months previously with a 1CU IOL prospectively randomly allocated to 1 eye and an AcrySof MA30 monofocal IOL to the other eye were examined. Distance correction, near vision, reading fluency, near point, and defocus to minus spheres were measured. Intraocular lens shift to an accommodative stimulus following instillation of pilocarpine 4% was measured with the ACMaster. Refractive change between distance and near was measured with the Tracey wavefront analyzer.

RESULTS

Of the original 30 patients recruited, complete measurements could only be obtained for 20. There was no significant difference in near visual function with either IOL. A small anterior movement of the 1CU was seen with accommodation 0.010 mm +/- 0.028 (SD). After pilocarpine 4% instillation, a forward movement of 0.220 +/- 0.169 mm was seen with the 1CU compared to a backward movement of 0.028 +/- 0.095 with the MA30. There was no significant correlation between distance corrected near visual acuity and IOL movement. No change in spherical equivalent between distance and near was seen on wavefront analysis of either IOL.

CONCLUSIONS

Small forward movement of the 1CU IOL was seen with accommodation and increased following pilocarpine, compared to the posterior movement of the MA30 IOL. The amount of the IOL shift was not sufficient to provide useful near vision, but the difference suggests that the engineering concept behind the 1CU IOL is valid.

摘要

目的

客观测量1CU可调节人工晶状体(IOL)(HumanOptics公司)的移位和屈光变化,并将其与同一只眼的单焦点AcrySof MA30 IOL(爱尔康实验室)进行比较。

设置

英国伦敦圣托马斯医院眼科。

方法

对30例患者进行检查,这些患者在18至24个月前接受了双侧白内障手术,其中一只眼植入1CU IOL,另一只眼植入AcrySof MA30单焦点IOL,是前瞻性随机分配的。测量了远距矫正、近视力、阅读流畅度、近点以及负球镜散焦。使用ACMaster测量滴入4%毛果芸香碱后人工晶状体对调节刺激的移位。使用Tracey波前分析仪测量远距和近距之间的屈光变化。

结果

在最初招募的30例患者中,仅20例获得了完整测量数据。两种IOL的近视力功能无显著差异。1CU在调节时可见微小的向前移位,为0.010 mm±0.028(标准差)。滴入4%毛果芸香碱后,1CU向前移位0.220±0.169 mm,而MA30向后移位0.028±0.095 mm。远距矫正近视力与IOL移位之间无显著相关性。两种IOL的波前分析均未发现远距和近距之间的等效球镜度变化。

结论

与MA30 IOL的向后移位相比,1CU IOL在调节时可见微小的向前移位,滴入毛果芸香碱后移位增加。IOL移位量不足以提供有用的近视力,但这种差异表明1CU IOL背后的工程理念是有效的。