Use of 18F-FDG-PET in the diagnosis of endoprosthetic loosening of knee and hip implants.

机构信息

Department of Orthopaedic Surgery, Klinikum Großhadern, Ludwig-Maximilians-University Munich, Marchioninistrasse 15, Munich, Germany.

出版信息

Arch Orthop Trauma Surg. 2010 Oct;130(10):1231-8. doi: 10.1007/s00402-009-1000-z. Epub 2009 Nov 5.

Abstract

INTRODUCTION

To diagnose septic and aseptic loosening 18-fluorodeoxyglucose positron emission tomography ((18)F-FDG-PET) has been described with good results for hip arthroplasties. The purpose of the present study was to examine whether there is a difference of feasibility in detecting loosening of hip versus knee prostheses by use of (18)F-FDG-PET.

PATIENTS

Thirty-two patients with lower limb arthroplasty complaints (74 components) were studied preoperatively with (18)F-FDG-PET. The interpretation of (18)F-FDG-PET was done according to evaluated criteria. The final diagnosis based on intraoperative findings in all cases including microbiological examinations.

RESULTS

For hip arthroplasty sensitivity/specificity of (18)F-FDG-PET towards implant loosening was 80%/87%. For infectious loosening of hip endoprostheses sensitivity/specificity was 67%/83%. In knee endoprostheses sensitivity/specificity for loosening was 56%/82% and 14%/89% for infection. The sensitivity of the results for knee and hip joints in regard to infectious versus aseptic loosening was significantly different.

CONCLUSION

We confirm that (18)F-FDG-PET is an appropriate tool to diagnose hip arthroplasty loosening. Differing from that (18)F-FDG-PET showed a significant lower sensitivity/specificity in detecting septic loosening of knee endoprostheses. It may therefore be necessary to use different methods to diagnose loosening of endoprostheses depending on the type of implant which is examined.

摘要

简介

正电子发射断层扫描(PET)已被描述为一种用于诊断髋关节置换术后感染性和无菌性松动的方法,其结果良好。本研究的目的是检查使用 18-氟脱氧葡萄糖(18F-FDG)PET 检测髋关节与膝关节假体松动的可行性是否存在差异。

患者

32 例下肢关节置换术后患者(74 个关节)在术前接受了 18F-FDG-PET 检查。18F-FDG-PET 的解读是根据评估标准进行的。最终诊断基于所有病例的术中发现,包括微生物学检查。

结果

对于髋关节置换术,18F-FDG-PET 对假体松动的敏感性/特异性为 80%/87%。对于髋关节假体感染性松动,敏感性/特异性为 67%/83%。在膝关节假体中,松动的敏感性/特异性分别为 56%/82%,感染的敏感性/特异性分别为 14%/89%。膝关节和髋关节的结果在感染性和无菌性松动方面的敏感性差异具有统计学意义。

结论

我们证实 18F-FDG-PET 是一种诊断髋关节置换术后松动的合适工具。与髋关节不同,18F-FDG-PET 检测膝关节假体感染性松动的敏感性/特异性显著降低。因此,根据所检查的植入物类型,可能需要使用不同的方法来诊断假体松动。