评估欧姆龙 HEM-907XL 示波法血压测量仪在全国性调查环境中的有效性。

Assessing the validity of the Omron HEM-907XL oscillometric blood pressure measurement device in a National Survey environment.

机构信息

Division of Health and Nutrition Examination Surveys, National Center for Health Statistics, Centers for Disease Control and Prevention, Hyattsville, MD 20782, USA.

出版信息

J Clin Hypertens (Greenwich). 2010 Jan;12(1):22-8. doi: 10.1111/j.1751-7176.2009.00199.x.

Abstract

Blood pressure (BP) readings taken by Omron HEM-907XL were compared with the results obtained using sphygmomanometer (HgS) in 509 individuals using 2002 Association for the Advancement of Medical Instrumentation (AAMI) criteria. With the exception of diastolic BP in youth ages 13 to 19 years (mean difference, -1.77 mm Hg; standard deviation, 8.65), the Omron device met the criteria. Agreement for hypertension (BP >or=140/90 mm Hg) was above chance (kappa=0.68) and, compared with HgS, Omron underestimated the prevalence of hypertension by 2.65%. The Omron and HgS measurements were highly correlated (r=0.94 for systolic BP and r=0.83 for diastolic BP). Both increased systolic and diastolic BP decreased device agreement (beta-coefficient=-0.10872, P<.0001; beta-coefficient=-0.25981, P<.0001, respectively). The Omron device meets AAMI criteria with the exception of diastolic BP in youth ages 13 to 19 years. However, Omron underestimated the prevalence of hypertension and device agreement decreases with increased systolic and diastolic BP.

摘要

使用 2002 年美国医疗器械促进协会(AAMI)标准,将欧姆龙 HEM-907XL 测量的血压(BP)读数与使用血压计(HgS)获得的 509 个人的结果进行了比较。除了 13 至 19 岁青少年的舒张压(平均差异,-1.77mmHg;标准差,8.65mmHg)外,欧姆龙设备符合标准。高血压(BP≥140/90mmHg)的一致性高于偶然机会(kappa=0.68),与 HgS 相比,欧姆龙低估了高血压的患病率 2.65%。欧姆龙和 HgS 测量结果高度相关(收缩压 r=0.94,舒张压 r=0.83)。收缩压和舒张压的升高都会降低设备的一致性(β系数=-0.10872,P<.0001;β系数=-0.25981,P<.0001)。欧姆龙设备符合 AAMI 标准,但 13 至 19 岁青少年的舒张压除外。然而,欧姆龙低估了高血压的患病率,并且随着收缩压和舒张压的升高,设备的一致性也会降低。

相似文献

1
Assessing the validity of the Omron HEM-907XL oscillometric blood pressure measurement device in a National Survey environment.
J Clin Hypertens (Greenwich). 2010 Jan;12(1):22-8. doi: 10.1111/j.1751-7176.2009.00199.x.
4
Assessing the validity of oscillometric device for blood pressure measurement in a large population-based epidemiologic study.
J Am Soc Hypertens. 2017 Nov;11(11):730-736.e4. doi: 10.1016/j.jash.2017.09.004. Epub 2017 Sep 21.
10
Accuracy of the Omron HEM-705 CP for blood pressure measurement in large epidemiologic studies.
Arq Bras Cardiol. 2011 May;96(5):393-8. doi: 10.1590/s0066-782x2011005000038. Epub 2011 Apr 1.

引用本文的文献

1
Sex Differences in BP From an Automated Oscillometric Compared With Manual Device.
Hypertension. 2025 Sep;82(9):1449-1459. doi: 10.1161/HYPERTENSIONAHA.124.24640. Epub 2025 Jul 1.
5
The Effect of Acute Sleep Extension on Blood Pressure Is Dependent on the Change in Sleep Efficiency.
Clocks Sleep. 2024 Oct 4;6(4):546-556. doi: 10.3390/clockssleep6040036.
10
Community health worker navigation to improve allostatic load: The Integrated Population Health (IPOP) study.
Contemp Clin Trials Commun. 2023 Nov 16;36:101235. doi: 10.1016/j.conctc.2023.101235. eCollection 2023 Dec.

本文引用的文献

1
Comparison between an automated and manual sphygmomanometer in a population survey.
Am J Hypertens. 2008 Mar;21(3):280-3. doi: 10.1038/ajh.2007.54. Epub 2008 Jan 24.
5
US demographic trends in mid-arm circumference and recommended blood pressure cuffs: 1988-2002.
J Hum Hypertens. 2005 Nov;19(11):885-91. doi: 10.1038/sj.jhh.1001905.
8
Measuring blood pressure accurately: new and persistent challenges.
JAMA. 2003 Feb 26;289(8):1027-30. doi: 10.1001/jama.289.8.1027.
9
Validation of the Omron HEM-907 device for blood pressure measurement.
Blood Press Monit. 2002 Aug;7(4):237-41. doi: 10.1097/00126097-200208000-00006.
10
Evaluation of the overall efficacy of the Omron office digital blood pressure HEM-907 monitor in adults.
Blood Press Monit. 2001 Apr;6(2):107-10. doi: 10.1097/00126097-200104000-00007.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验