Division of Cardiovascular Surgery, Peter Munk Cardiac Centre at Toronto General Hospital, Toronto, Ontario, Canada.
J Thorac Cardiovasc Surg. 2012 Apr;143(4 Suppl):S74-7. doi: 10.1016/j.jtcvs.2011.10.030. Epub 2011 Nov 20.
This study was undertaken to examine the possible adverse effect of the mitral valve prosthesis on the hemodynamic performance of the aortic valve prosthesis in patients who have undergone double valve replacement.
Patients who underwent double valve replacement were matched for age, body surface area, left ventricular function, and size and type of aortic valve prosthesis with patients who underwent isolated aortic valve replacement. Two types of prosthetic valves were examined: the St Jude Medical mechanical valve (St Jude Medical, St Paul, Minn) and the Hancock II bioprosthesis (Medtronic Inc, Minneapolis, Minn). Five patients for each size and type of aortic valve prosthesis in the double valve replacement group were matched at 1:2 with patients in the isolated aortic valve replacement group. Only valve sizes 21 to 27 were matched. Hemodynamic assessment of the aortic valve prosthesis was performed by transthoracic echocardiogram before hospital discharge.
Matched patients had similar clinical profiles. There were no differences in the systolic gradients, effective aortic valve areas, or flow velocity across the aortic valve prostheses after isolated aortic valve replacement or double valve replacement.
Early after surgery, the hemodynamic performance of aortic valve prostheses was not affected by the presence of mitral valve prostheses in patients who underwent combined aortic and mitral valve replacement.
本研究旨在探讨二尖瓣假体对行双瓣置换术患者主动脉瓣假体血流动力学性能的可能不良影响。
将行双瓣置换术的患者与行单纯主动脉瓣置换术的患者进行年龄、体表面积、左心室功能、主动脉瓣假体大小和类型的匹配。检查了两种类型的假体瓣膜:St Jude Medical 机械瓣膜(明尼苏达州圣保罗的 St Jude Medical)和 Hancock II 生物瓣膜(明尼苏达州明尼阿波利斯的 Medtronic Inc)。双瓣置换组中每个主动脉瓣假体的 5 名患者与单纯主动脉瓣置换组中相应的 2:1 患者相匹配。仅匹配瓣膜尺寸为 21 至 27 的患者。在出院前通过经胸超声心动图对主动脉瓣假体进行血流动力学评估。
匹配患者的临床特征相似。在单纯主动脉瓣置换或双瓣置换后,主动脉瓣假体的收缩期梯度、有效主动脉瓣面积或跨主动脉瓣假体的流速均无差异。
手术后早期,行主动脉瓣和二尖瓣联合置换术的患者,二尖瓣假体的存在并未影响主动脉瓣假体的血流动力学性能。