Department of Pediatrics, Seoul National University College of Medicine, Seoul, Republic of Korea.
Pediatr Crit Care Med. 2013 Jan;14(1):70-3. doi: 10.1097/PCC.0b013e318260117d.
To evaluate clinical applicability of noninvasive hemoglobin (Hb) measurement with a pulse CO-oximeter in neonates.
Prospective comparison study.
Neonatal ICU.
Fifty-six preterm and term infants with median age = 20 days (range = 1-98 days) and median weight = 1,440 g (range = 530-4,230 g).
Hb measurements by Pulse CO-Oximetry (Masimo Radical-7) were recorded immediately prior to venous samplings.
The collected data were compared with the corresponding venous Hb level obtained in laboratory testing, and a total of 137 data pairs were analyzed. Noninvasive Hb values measured with a pulse CO-oximeter were significantly correlated with the venous Hb levels (correlation coefficient, r = 0.758; p < 0.001). Hb values measured with a pulse CO-oximeter were higher than those measured with a laboratory hematology analyzer (13.3 ± 2.6g/dL vs. 12.5 ± 3.1g/dL). In terms of the agreement between the laboratory analyzer and the pulse CO-oximeter, 94.8% of the measurements fell within two standard deviations of the mean difference.
Noninvasive Hb measurements with Pulse CO-Oximetry provide clinically acceptable accuracy, and they were significantly correlated with laboratory Hb measurement in neonates. In terms of the clinical applicability, noninvasive Hb monitoring with a pulse CO-oximeter could be useful in the early detection of Hb changes in neonates.
评估脉搏 CO- 氧饱和度仪无创血红蛋白(Hb)测量在新生儿中的临床适用性。
前瞻性比较研究。
新生儿重症监护病房。
56 例早产儿和足月儿,中位年龄=20 天(范围=1-98 天),中位体重=1440g(范围=530-4230g)。
在进行静脉取样前,立即使用脉搏 CO- 血氧计(Masimo Radical-7)记录 Hb 测量值。
将收集的数据与实验室检测获得的相应静脉 Hb 水平进行比较,共分析了 137 对数据。脉搏 CO- 氧饱和度仪无创测量的 Hb 值与静脉 Hb 水平显著相关(相关系数,r=0.758;p<0.001)。脉搏 CO- 氧饱和度仪测量的 Hb 值高于实验室血液分析仪(13.3±2.6g/dL 与 12.5±3.1g/dL)。就实验室分析仪和脉搏 CO- 氧饱和度仪之间的一致性而言,94.8%的测量值落在平均差值的两个标准差内。
脉搏 CO- 氧饱和度仪的无创 Hb 测量具有临床可接受的准确性,与新生儿实验室 Hb 测量显著相关。就临床适用性而言,脉搏 CO- 氧饱和度仪的无创 Hb 监测可用于早期发现新生儿 Hb 变化。