Suppr超能文献

用于细胞外基质支架支持的人诱导多能干细胞衍生视网膜色素上皮细胞片移植的手术装置和手术程序的评估

Evaluation of the Surgical Device and Procedure for Extracellular Matrix-Scaffold-Supported Human iPSC-Derived Retinal Pigment Epithelium Cell Sheet Transplantation.

作者信息

Kamao Hiroyuki, Mandai Michiko, Ohashi Wataru, Hirami Yasuhiko, Kurimoto Yasuo, Kiryu Junichi, Takahashi Masayo

机构信息

Laboratory for Retinal Regeneration, RIKEN Center for Developmental Biology, Chuo-ku, Kobe, Hyogo, Japan 2Department of Ophthalmology, Kawasaki Medical School, Kurashiki, Okayama, Japan.

Laboratory for Retinal Regeneration, RIKEN Center for Developmental Biology, Chuo-ku, Kobe, Hyogo, Japan.

出版信息

Invest Ophthalmol Vis Sci. 2017 Jan 1;58(1):211-220. doi: 10.1167/iovs.16-19778.

Abstract

PURPOSE

To develop a clinically applicable transplantation device and surgical procedure for extracellular matrix-scaffold-supported human-induced pluripotent stem cell-derived retinal pigment epithelium (hiPSC-RPE) cell sheet transplantation for clinical use.

METHODS

The developed surgical device consisted of a custom-designed hand piece and a cannula. The subretinal transplantation of hiPSC-RPE cell sheets was performed in 12 rabbits. The results evaluated were the graft condition (damage or fold), side (front or back), position (center, near, far), and direction (anterior, posterior, right, left) immediately after surgery and the graft condition (shrinking or fold) 2 weeks after surgery. These results were evaluated by fundus photography and optical coherence tomography, followed by immersion-fixed histology.

RESULTS

All grafts could be transplanted without obvious damage. The transplanted grafts included 2 of 12 folded grafts, 12 of 12 front side, 12 of 12 center position, 10 of 12 anterior direction, and 2 of 12 right direction immediately after surgery, whereas transplantation with a distance between an inlet and an outlet greater than graft and the coaxial direction of the flow paths and the insertion device posed the correct condition and direction. Two weeks after the surgery, the transplanted grafts included two folded grafts and four shrunken grafts; however, complete drainage of subretinal fluid for adhesion between the graft and the host prevented shrunken grafts.

CONCLUSIONS

A developed surgical device and procedure allow grafts to be transplanted into the targeted transplantation site safely and reproducibly. This surgical method will provide additional information on the advancement of future RPE transplantation therapies.

摘要

目的

开发一种临床适用的移植装置和手术方法,用于细胞外基质支架支持的人诱导多能干细胞衍生的视网膜色素上皮(hiPSC-RPE)细胞片移植以供临床使用。

方法

所开发的手术装置由定制设计的手持件和套管组成。在12只兔子中进行hiPSC-RPE细胞片的视网膜下移植。评估的结果包括术后即刻的移植物状况(损伤或褶皱)、面(正面或背面)、位置(中央、近处、远处)和方向(前部、后部、右侧、左侧),以及术后2周的移植物状况(收缩或褶皱)。通过眼底摄影和光学相干断层扫描评估这些结果,随后进行浸入固定组织学检查。

结果

所有移植物均可在无明显损伤的情况下进行移植。术后即刻,移植的移植物中12个有2个褶皱,12个为正面,12个位于中央位置,12个中有10个方向为前部,12个中有2个方向为右侧,而入口与出口之间的距离大于移植物且流动路径与插入装置同轴方向时移植条件和方向正确。术后2周,移植的移植物中有2个褶皱移植物和4个收缩移植物;然而,视网膜下液的完全引流以防止移植物与宿主之间粘连可防止移植物收缩。

结论

所开发的手术装置和方法可使移植物安全且可重复地移植到目标移植部位。这种手术方法将为未来RPE移植治疗的进展提供更多信息。