Simpson Joanna P, Fascione Martin, Bergström Ed, Wilson Julie, Collins Matthew J, Penkman Kirsty E H, Thomas-Oates Jane
Department of Chemistry, University of York, York, UK.
Centre of Excellence in Mass Spectrometry, University of York, York, UK.
Rapid Commun Mass Spectrom. 2019 Jun 30;33(12):1049-1057. doi: 10.1002/rcm.8441.
Although mass spectrometry (MS) is routinely used to determine deamination in peptide mixtures, the effects of the choice of ionisation source have not yet been investigated. In particular, matrix-assisted laser desorption/ionisation (MALDI) has become a popular tool with which to measure levels of glutamine deamidation in ancient proteins. Here we use model synthetic peptides to rigorously compare MALDI and electrospray ionisation (ESI).
We used two synthetic peptides, with glutamine (Q) in one substituted for glutamic acid (E) in the other, to investigate the suitability of MALDI and ESI sources for the assessment of deamidation in peptides using MS. We also compared measurements of the same Q- and E-containing peptide mixtures using two different mass analysers (time-of-flight (TOF) and Fourier transform ion cyclotron resonance (FT-ICR)).
When standard mixtures of the Q- and E-containing peptides were analysed using MALDI, under-representation of the E-containing peptide was observed. This observation was consistent between analyses carried out using either TOF or FT-ICR-MS. When the same mixtures were analysed using ESI FT-ICR-MS, no ionisation bias was observed.
MALDI may not be a suitable ionisation method for the determination of deamidation in peptide mixtures. However, ESI was successfully used to determine the ratio in known mixtures of Q- and E-containing peptides. These preliminary observations warrant further investigation into ionisation bias when measuring deamidation in other peptide sequences.
尽管质谱法(MS)通常用于测定肽混合物中的脱氨基作用,但尚未研究电离源选择的影响。特别是,基质辅助激光解吸/电离(MALDI)已成为测量古代蛋白质中谷氨酰胺脱酰胺水平的常用工具。在此,我们使用模型合成肽来严格比较MALDI和电喷雾电离(ESI)。
我们使用了两种合成肽,一种含有谷氨酰胺(Q),另一种用谷氨酸(E)替代,以研究MALDI和ESI源用于通过MS评估肽中脱酰胺作用的适用性。我们还比较了使用两种不同质量分析器(飞行时间(TOF)和傅里叶变换离子回旋共振(FT-ICR))对相同含Q和含E肽混合物的测量结果。
当使用MALDI分析含Q和含E肽的标准混合物时,观察到含E肽的代表性不足。使用TOF或FT-ICR-MS进行的分析之间,这一观察结果是一致的。当使用ESI FT-ICR-MS分析相同混合物时,未观察到电离偏差。
MALDI可能不是测定肽混合物中脱酰胺作用的合适电离方法。然而,ESI成功用于测定含Q和含E肽的已知混合物中的比例。这些初步观察结果值得进一步研究在测量其他肽序列中的脱酰胺作用时的电离偏差。