From the Departments of Imaging Physics (M.C.J., D.D.C., R.R.L.), Interventional Radiology (E.N.K.C., D.L.B.), Diagnostic Radiology, Abdominal Imaging Section (E.P.T.), and Diagnostic Radiology, Neuroradiology Section (D.S.), The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd, Houston, TX 77030; and Radiology, Medical Physics Section (X.D.), University of Texas Southwestern Medical Center, Dallas, Tex.
Radiology. 2019 Aug;292(2):414-419. doi: 10.1148/radiol.2019182870. Epub 2019 Jun 25.
Background Assessments of the quantitative limitations among the six commercially available dual-energy (DE) CT acquisition schemes used by major CT manufacturers could aid researchers looking to use iodine quantification as an imaging biomarker. Purpose To determine the limits of detection and quantification of DE CT in phantoms by comparing rapid peak kilovoltage switching, dual-source, split-filter, and dual-layer detector systems in six different scanners. Materials and Methods Seven 50-mL iohexol solutions were used, with concentrations of 0.03-2.0 mg iodine per milliliter. The solutions and water sample were scanned five times each in two phantoms (small, 20-cm diameter; large, 30 × 40-cm diameter) with six DE CT systems and a total of 10 peak kilovoltage settings or combinations. Iodine maps were created, and the mean iodine signal in each sample was recorded. The limit of blank (LOB) was defined as the upper limit of the 95% confidence interval of the water sample. The limit of detection (LOD) was defined as the concentration with a 95% chance of having a signal above the LOB. The limit of quantification (LOQ) was defined as the lowest concentration where the coefficient of variation was less than 20%. Results The LOD range was 0.021-0.26 mg/mL in the small phantom and 0.026-0.55 mg/mL in the large phantom. The LOQ range was 0.07-0.50 mg/mL in the small phantom and 0.20-1.0 mg/mL in the large phantom. The dual-source and rapid peak kilovoltage switching systems had the lowest LODs, and the dual-layer detector systems had the highest LODs. Conclusion The iodine limit of detection using dual-energy CT systems varied with scanner and phantom size, but all systems depicted iodine in the small and large phantoms at or below 0.3 and 0.5 mg/mL, respectively, and enabled quantification at concentrations of 0.5 and 1.0 mg/mL, respectively. © RSNA, 2019 See also the editorial by Hindman in this issue.
背景 对主要 CT 制造商使用的六种商业双能 (DE) CT 采集方案的定量限制进行评估,可以帮助研究人员将碘定量作为成像生物标志物。目的 比较快速峰值千伏切换、双源、分束滤光片和双层探测器系统,在六台不同的扫描仪中确定 DE CT 在体模中的检测限和定量极限。材料与方法 使用了 7 种 50 毫升的碘海醇溶液,浓度为每毫升 0.03-2.0 毫克碘。在两个体模(小体模直径 20 厘米;大体模 30×40 厘米)中,使用六种 DE CT 系统和总共 10 种峰值千伏设置或组合,对每个溶液和水样本进行五次扫描。创建碘图,并记录每个样本中的平均碘信号。将空白上限 (LOB) 定义为水样本 95%置信区间上限。检测限 (LOD) 定义为信号超过 LOB 的概率为 95%时的浓度。定量限 (LOQ) 定义为变异系数小于 20%时的最低浓度。结果 在小体模中,LOD 范围为 0.021-0.26 毫克/毫升,在大体模中,LOD 范围为 0.026-0.55 毫克/毫升。在小体模中,LOQ 范围为 0.07-0.50 毫克/毫升,在大体模中,LOQ 范围为 0.20-1.0 毫克/毫升。双源和快速峰值千伏切换系统的 LOD 最低,双层探测器系统的 LOD 最高。结论 使用 DE CT 系统的碘检测限随扫描仪和体模大小而变化,但所有系统在小和大体模中均能检测到低于 0.3 和 0.5 毫克/毫升的碘,并分别在 0.5 和 1.0 毫克/毫升的浓度下实现定量。