Classical posturography techniques have been recently enhanced by the use of different motion tracking devices, but for technical reasons they are not used to track directly the body spatial position of a subject.
To describe and clinically evaluate a wireless inertial measurement unit-based mobile system to track body position changes.
The developed system used a calculus transformation method using the acceleration data corrected by Kalman and Butterworth filters to output position data. A prospective non-randomized clinical study involving 15 healthy subjects was performed to evaluate the agreement between the confidence ellipse areas synchronously measured by the new developed system and a classical posturography system while performing a modified clinical test of sensory interaction in balance.
The overall intra-class correlation index was 0.93 (CI 0.89, 0.96). Grouped by conditions, under conditions 1-4, Pearson's correlation was 0.604, 0.78, 0.882, and 0.81, respectively.
The developed wireless inertial measurement unit-based posturography system was valid for tracking the sway variances in normal subjects under habitual clinical testing conditions. Further studies are needed to validate this system on patients and also under other posture conditions.
经典的姿势描记术技术最近通过使用不同的运动跟踪设备得到了增强,但由于技术原因,它们不能直接用于跟踪受试者的身体空间位置。
描述和临床评估一种基于无线惯性测量单元的移动系统,以跟踪身体位置变化。
开发的系统使用了一种微积分变换方法,使用卡尔曼和巴特沃斯滤波器校正的加速度数据来输出位置数据。进行了一项前瞻性的非随机临床研究,涉及 15 名健康受试者,以评估在进行改良平衡感觉相互作用临床测试时,新开发系统同步测量的置信椭圆区域与经典姿势描记系统之间的一致性。
总体组内相关指数为 0.93(CI 0.89, 0.96)。按条件分组,条件 1-4 下,皮尔逊相关系数分别为 0.604、0.78、0.882 和 0.81。
开发的基于无线惯性测量单元的姿势描记系统可用于在常规临床测试条件下跟踪正常受试者的摆动方差。需要进一步的研究来验证该系统在患者和其他姿势条件下的有效性。