Graduate School of Science and Technology, Tokushima University, Tokushima, 770-8506, Japan.
Graduate School of Science and Technology, Tokushima University, Tokushima, 770-8506, Japan.
Anal Chim Acta. 2020 Mar 15;1102:46-52. doi: 10.1016/j.aca.2019.12.045. Epub 2019 Dec 18.
The electrochemical flow cell containing track-etched microporous membrane electrodes was applied to a dual-electrode coulometric detector for microbore/capillary HPLC with a small injection volume and low eluent flow rate. The proposed flow cell with a 0.1-mm diameter inlet channel gave a detection volume of 0.08 nL per electrode, which was determined by the eluent flow through the electrode. For the dual-electrode detector, the calculated volume was 0.24 nL. The efficiency of electrooxidation of l-ascorbic acid increased as the flow rate decreased and was close to 100% when the flow rate was below 50 μL min, which is a common flow rate in microbore or capillary liquid chromatography. Catecholamines, such as noradrenaline, adrenaline, and dopamine, were detected by total conversion with two-electron oxidation in the potential range from 0.8 to 1.0 V vs. Ag/AgCl after separation with a microbore column. These peaks were accompanied by corresponding cathodic peaks derived from quasi-stable electrooxidation products of the catecholamines. The detection limits of noradrenaline, adrenaline, and dopamine were 0.1, 0.1, and 0.2 μM, respectively. The RSD values for five replicate measurements of 5.0 μM of these compounds were 0.9%, 0.7%, and 1.5%, respectively. Coulometric detection was also demonstrated by determination of catecholamines in pharmaceuticals.
电化学流通池包含刻蚀微孔膜电极,被应用于双电极库仑计检测器,用于微径柱/毛细管 HPLC,具有小进样体积和低洗脱液流速。该流通池的进口通道直径为 0.1 毫米,每个电极的检测体积为 0.08 nL,这是由通过电极的洗脱液决定的。对于双电极检测器,计算的体积为 0.24 nL。当流速低于 50 μL/min 时,抗坏血酸的电氧化效率随着流速的降低而增加,接近 100%,这是微径柱或毛细管液相色谱中的常见流速。儿茶酚胺,如去甲肾上腺素、肾上腺素和多巴胺,在微径柱分离后,在 0.8 到 1.0 V 相对于 Ag/AgCl 的电位范围内,通过双电子氧化被完全转化为电信号而被检测到。这些峰伴随着源自儿茶酚胺的准稳定电氧化产物的相应阴极峰。去甲肾上腺素、肾上腺素和多巴胺的检测限分别为 0.1、0.1 和 0.2 μM。这三种化合物的 5.0 μM 五个重复测量的 RSD 值分别为 0.9%、0.7%和 1.5%。通过测定药物中的儿茶酚胺也证明了库仑检测的有效性。