Kestrel Labs, Inc., Research and Development, Boulder, CO, USA.
Division of Emergency Medicine, Department of Pediatrics, Vanderbilt University School of Medicine, Vanderbilt Children's Hospital, Nashville, TN, USA.
J Clin Monit Comput. 2022 Apr;36(2):579-586. doi: 10.1007/s10877-021-00704-1. Epub 2021 Apr 19.
Almost since its introduction pulse oximetry was known to overestimate oxygen saturation in cases of carbon monoxide poisoning or elevated methemoglobin (metHb) levels. To eliminate this dangerous behavior some manufacturers have added additional LED emitters to try to increase the number of measured hemoglobin species and to improve measurement accuracy, but have not been very successful. We hypothesized that the use of narrow-band laser light sources would make accurate and precise measurement of the four primary species of hemoglobin possible, even in cases of elevated levels of carboxyhemoglobin (COHb). Calibration and verification studies were performed on a tissue simulator that employed an artificial finger pulsating with whole human blood. This simulator allowed safe generation of 165 different combinations of the levels of oxyhemoglobin (OHb), COHb, metHb, and reduced hemoglobin (RHb) for calibration of the laser-based pulse oximeter. A follow-on study used 56 mixed hemoglobin levels for verification and statistical analysis of the performance of this device. This laser-based pulse oximeter measured all four species of hemoglobin accurately and precisely (A ≤ 1.8%) for metHb levels in the clinically normal range. At elevated metHb levels the device continued to provide accurate and precise measurements of metHb and RHb (A ≤ 1.7%). The use of monochromatic laser light sources can create a new generation of highly accurate, multi-parameter, pulse oximeters.
几乎从引入脉搏血氧仪开始,人们就知道在一氧化碳中毒或高铁血红蛋白(metHb)水平升高的情况下,它会高估氧饱和度。为了消除这种危险行为,一些制造商增加了额外的 LED 发射器,试图增加测量的血红蛋白种类数量并提高测量精度,但效果并不理想。我们假设使用窄带激光光源可以实现对四种主要血红蛋白的准确和精确测量,即使在碳氧血红蛋白(COHb)水平升高的情况下也是如此。我们在一个使用全血搏动的人工手指的组织模拟器上进行了校准和验证研究。该模拟器允许安全地生成 165 种不同的氧合血红蛋白(OHb)、COHb、metHb 和还原血红蛋白(RHb)水平组合,用于校准基于激光的脉搏血氧仪。后续研究使用 56 种混合血红蛋白水平对该设备的性能进行了验证和统计分析。这种基于激光的脉搏血氧仪可以准确、精确地测量所有四种血红蛋白(A ≤ 1.8%),用于临床正常范围内的 metHb 水平。在 metHb 水平升高的情况下,该设备继续提供准确和精确的 metHb 和 RHb 测量(A ≤ 1.7%)。使用单色激光光源可以创造新一代高度精确、多参数的脉搏血氧仪。