Real-World Use of a New Hybrid Closed Loop Improves Glycemic Control in Youth with Type 1 Diabetes.

机构信息

Barbara Davis Center for Diabetes, University of Colorado Anschutz, Aurora, Colorado, USA.

Diabetes Institute, University of Florida, Gainesville, Florida, USA.

出版信息

Diabetes Technol Ther. 2021 Dec;23(12):837-843. doi: 10.1089/dia.2021.0165. Epub 2021 Jun 21.

Abstract

To describe real-world outcomes for youth using the Tandem t:slim X2 insulin pump with Control-IQ technology ("Control-IQ") for 6 months at a large pediatric clinic. Youth with type 1 diabetes, who started Control-IQ for routine care, were prospectively followed. Data on system use and glycemic control were collected before Control-IQ start, and at 1, 3, and 6 months after start. Mixed models assessed change across time; interactions with baseline hemoglobin A1c (HbA1c) and age were tested. In 191 youth (median age 14, 47% female, and median HbA1c 7.6%), percent time with glucose levels 70-180 mg/dL (time-in-range [TIR]) improved from 57% at baseline to 66% at 6 months ( < 0.001). The proportion of participants reaching TIR target (>70%) doubled from 23.5% at baseline to 47.8% at 3 months, sustaining at 46.7% at 6 months ( < 0.001). Glucose management indicator (approximation of HbA1c) improved from 7.5% at baseline to 7.1% at 3 months and 7.2% at 6 months ( < 0.001). Those with higher baseline HbA1c experienced the most substantial improvements in glycemic control. Percent time using the Control-IQ feature was 86.4% at 6 months, and <4% of cohort discontinued use. The Control-IQ system clinically and significantly improved glycemic control in a large sample of youth. System use was high at 6 months, with only a small proportion discontinuing use, indicating potential for sustaining results long term.

摘要

描述在一家大型儿科诊所使用 Tandem t:slim X2 胰岛素泵和 Control-IQ 技术(“Control-IQ”)进行 6 个月的真实世界青少年结果。患有 1 型糖尿病的青少年开始常规使用 Control-IQ,前瞻性随访。在开始使用 Control-IQ 之前、开始后 1、3 和 6 个月收集系统使用和血糖控制数据。混合模型评估随时间的变化;测试与基线血红蛋白 A1c(HbA1c)和年龄的交互作用。在 191 名青少年(中位年龄 14 岁,47%为女性,中位 HbA1c 为 7.6%)中,血糖水平在 70-180mg/dL(时间在范围内[TIR])的时间百分比从基线的 57%提高到 6 个月时的 66%( < 0.001)。达到 TIR 目标(>70%)的参与者比例从基线时的 23.5%增加到 3 个月时的 47.8%,并在 6 个月时持续保持在 46.7%( < 0.001)。血糖管理指标(HbA1c 的近似值)从基线时的 7.5%改善到 3 个月时的 7.1%和 6 个月时的 7.2%( < 0.001)。那些基线时 HbA1c 较高的人血糖控制改善最显著。在 6 个月时,使用 Control-IQ 功能的时间百分比为 86.4%,不到 4%的患者停止使用。Control-IQ 系统在一个大样本的青少年中显著改善了血糖控制。在 6 个月时,系统使用率很高,只有少数人停止使用,表明有潜力长期维持治疗效果。

相似文献

1
Real-World Use of a New Hybrid Closed Loop Improves Glycemic Control in Youth with Type 1 Diabetes.
Diabetes Technol Ther. 2021 Dec;23(12):837-843. doi: 10.1089/dia.2021.0165. Epub 2021 Jun 21.
5
Six months of hybrid closed loop in the real-world: An evaluation of children and young adults using the 670G system.
Pediatr Diabetes. 2020 Mar;21(2):310-318. doi: 10.1111/pedi.12962. Epub 2020 Jan 7.
6
Glycemic Outcomes in Baseline Hemoglobin A1C Subgroups in the International Diabetes Closed-Loop Trial.
Diabetes Technol Ther. 2022 Aug;24(8):588-591. doi: 10.1089/dia.2021.0524. Epub 2022 Feb 8.
7
Six-Month Effectiveness of Advanced Standard Hybrid Closed-Loop System in Children and Adolescents With Type 1 Diabetes Mellitus.
Front Endocrinol (Lausanne). 2021 Nov 9;12:766314. doi: 10.3389/fendo.2021.766314. eCollection 2021.
8
One Year Real-World Use of the Control-IQ Advanced Hybrid Closed-Loop Technology.
Diabetes Technol Ther. 2021 Sep;23(9):601-608. doi: 10.1089/dia.2021.0097. Epub 2021 Apr 21.
9
Real-World Glycemic Outcomes with Early Omnipod 5 Use in Youth with Type 1 Diabetes.
Diabetes Technol Ther. 2023 Nov;25(11):782-789. doi: 10.1089/dia.2023.0337. Epub 2023 Sep 20.
10
Real-World Patient-Reported Outcomes and Glycemic Results with Initiation of Control-IQ Technology.
Diabetes Technol Ther. 2021 Feb;23(2):120-127. doi: 10.1089/dia.2020.0388. Epub 2020 Sep 10.

引用本文的文献

4
Real-World Glycemic Improvements with Hybrid Closed Loop Pumps in Youth with Type 1 Diabetes.
Pediatr Diabetes. 2023 Oct 5;2023:6621706. doi: 10.1155/2023/6621706. eCollection 2023.
7
14. Children and Adolescents: Standards of Care in Diabetes-2025.
Diabetes Care. 2025 Jan 1;48(Supplement_1):S283-S305. doi: 10.2337/dc25-S014.
10
Use of diabetes technology in children.
Diabetologia. 2024 Oct;67(10):2075-2084. doi: 10.1007/s00125-024-06218-0. Epub 2024 Jul 12.

本文引用的文献

1
Initiating hybrid closed loop: A program evaluation of an educator-led Control-IQ follow-up at a large pediatric clinic.
Pediatr Diabetes. 2021 Jun;22(4):586-593. doi: 10.1111/pedi.13183. Epub 2021 Mar 16.
2
13. Children and Adolescents: .
Diabetes Care. 2021 Jan;44(Suppl 1):S180-S199. doi: 10.2337/dc21-S013.
3
6. Glycemic Targets: .
Diabetes Care. 2021 Jan;44(Suppl 1):S73-S84. doi: 10.2337/dc21-S006.
4
A Randomized Trial of Closed-Loop Control in Children with Type 1 Diabetes.
N Engl J Med. 2020 Aug 27;383(9):836-845. doi: 10.1056/NEJMoa2004736.
6
Cost, Hassle, and On-Body Experience: Barriers to Diabetes Device Use in Adolescents and Potential Intervention Targets.
Diabetes Technol Ther. 2020 Oct;22(10):760-767. doi: 10.1089/dia.2019.0509. Epub 2020 Mar 27.
8
Six months of hybrid closed loop in the real-world: An evaluation of children and young adults using the 670G system.
Pediatr Diabetes. 2020 Mar;21(2):310-318. doi: 10.1111/pedi.12962. Epub 2020 Jan 7.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验