Abstract

The series of tests described in a preceding publication (Spicher and Borchers, 1983) has been continued in a modified way. This time, the dependency of the microbiological test results of a formaldehyde gas sterilization procedure on length and inner diameter of the tubes serving as test pieces was examined. The tubes were 1 or 2 m in length with an inner diameter of 1 or 2 mm. The tests were performed with four different preparations of bioindicators. Spores of Bac. stearothermophilus served as test germs. The preparations differed in the type of suspension used for the preparation of the bioindicators: distilled water, diluted blood (10%), undiluted blood, 10% albumin solution. The spore suspensions had been dried on linen thread. During the test procedure, the bioindicators were located near the sealed end of the tube. After completion of the sterilization procedure, the bioindicators were examined for viable germs. In tubes of identical length, the frequency of indicators carrying viable germs was always higher in those of 1 mm than in those of 2 mm inner diameter. In tubes of identical inner diameter, the frequency of indicators carrying viable germs in those of 2 m length was always higher than in those of 1 m length. This regularity was independent of the type of bioindicators used. The bioindicators for the preparation of which a 10% albumin solution had been employed showed the highest resistance. A somewhat lower resistance was found for the bioindicators prepared with undiluted blood. The bioindicators for which the spores had been suspended in diluted blood proved to have the lowest resistance. If the spores had been suspended in distilled water, the resistance of the bioindicators was a little lower than that of those suspended in undiluted blood, but was higher than that of the dried spores with diluted blood. The test results confirm the effectiveness of the method proposed earlier, i.e. to deposit the bioindicators in special test pieces (e.g. tubes or sounds) for the microbiological testing of formaldehyde gas sterilization procedures. These test pieces must be at least as long and as narrow as the longest and narrowest cavity of the object to be sterilized (tubes, catheters). In order to standardize the microbiological testing of formaldehyde gas sterilization procedures and to guarantee a certain minimum efficiency, the bioindicator as well as the test piece and its size (length and inner diameter) should be standardized.(ABSTRACT TRUNCATED AT 400 WORDS)

摘要

前一篇出版物(斯皮彻和博彻斯,1983年)中描述的一系列测试以改进的方式继续进行。这次,研究了甲醛气体灭菌程序的微生物测试结果对用作测试样品的管子长度和内径的依赖性。管子长度为1米或2米,内径为1毫米或2毫米。测试使用了四种不同制备方法的生物指示剂。嗜热脂肪芽孢杆菌的孢子用作测试菌。这些制剂在用于制备生物指示剂的悬浮液类型上有所不同:蒸馏水、稀释血液(10%)、未稀释血液、10%白蛋白溶液。孢子悬浮液已干燥在亚麻线上。在测试过程中,生物指示剂位于管子的密封端附近。灭菌程序完成后,检查生物指示剂是否有活的细菌。在长度相同的管子中,内径为1毫米的管子中携带活细菌的指示剂频率总是高于内径为2毫米的管子。在内径相同的管子中,长度为2米的管子中携带活细菌的指示剂频率总是高于长度为1米的管子。这种规律与所使用的生物指示剂类型无关。使用10%白蛋白溶液制备的生物指示剂显示出最高的抗性。用未稀释血液制备的生物指示剂抗性略低。孢子悬浮在稀释血液中的生物指示剂抗性最低。如果孢子悬浮在蒸馏水中,生物指示剂的抗性比悬浮在未稀释血液中的略低,但高于用稀释血液干燥的孢子。测试结果证实了先前提出的方法的有效性,即将生物指示剂放置在特殊的测试样品(如管子或探子)中用于甲醛气体灭菌程序的微生物测试。这些测试样品必须至少与待灭菌物体(管子、导管)最长和最窄的腔一样长和窄。为了使甲醛气体灭菌程序的微生物测试标准化并保证一定的最低效率,生物指示剂以及测试样品及其尺寸(长度和内径)都应标准化。(摘要截选至400字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验