一项关于深部经颅磁刺激的随机对照可行性与安全性研究。

A randomized controlled feasibility and safety study of deep transcranial magnetic stimulation.

作者信息

Levkovitz Yechiel, Roth Yiftach, Harel Eiran Vadim, Braw Yoram, Sheer Aharon, Zangen Abraham

机构信息

Cognitive and Emotional Laboratory, Shalvata Mental Health Care Center, Hod-Hasharon, Israel.

出版信息

Clin Neurophysiol. 2007 Dec;118(12):2730-44. doi: 10.1016/j.clinph.2007.09.061. Epub 2007 Oct 30.

Abstract

OBJECTIVE

The H-coils are a new development in transcranial magnetic stimulation (TMS) research, allowing direct stimulation of deeper neuronal pathways than does standard TMS. This study assessed possible health risks, and some cognitive and emotional effects, of two H-coil versions designed to stimulate deep portions of the prefrontal cortex, using several stimulation frequencies.

METHODS

Healthy volunteers (n=32) were randomly assigned to one of four groups: each of two H-coil designs (H1/H2), standard figure-8 coil, and sham-coil control. Subjects were tested in a pre-post design, during three increasing (single pulses, 10 Hz, and 20 Hz) stimulation sessions, as well as 24-36 h after the last stimulation.

RESULTS

The major finding of the present study is that stimulation with the novel H-coils was well tolerated, with no adverse physical or neurological outcomes. Computerized cognitive tests found no deterioration in cognitive functions, except for a transient short-term effect of the H1-coil on spatial recognition memory on the first day of rTMS (but not in the following treatment days). On the other hand, spatial working memory was transiently improved by the H2-coil treatment. Finally, the questionnaires showed no significant emotional or mood alterations, except for reports on 'detachment' experienced by subjects treated with the H1-coil.

CONCLUSIONS

This study provides additional evidence for the feasibility and safety of the two H-coil designs (H1/H2).

SIGNIFICANCE

The H-coils offer a safe new tool with potential for both research and clinical applications for psychiatric and neurological disorders associated with dysfunctions of deep brain regions.

摘要

目的

H 型线圈是经颅磁刺激(TMS)研究中的一项新进展,与标准 TMS 相比,它能够直接刺激更深层的神经通路。本研究评估了两种旨在刺激前额叶皮质深部的 H 型线圈在使用几种刺激频率时可能存在的健康风险以及一些认知和情绪影响。

方法

健康志愿者(n = 32)被随机分配到四组中的一组:两种 H 型线圈设计(H1/H2)中的每一种、标准的 8 字线圈以及假线圈对照组。受试者采用前后测试设计,在三个递增的(单脉冲、10Hz 和 20Hz)刺激阶段进行测试,以及在最后一次刺激后 24 - 36 小时进行测试。

结果

本研究的主要发现是,新型 H 型线圈刺激耐受性良好,没有不良的身体或神经学后果。计算机化认知测试发现认知功能没有恶化,除了在重复经颅磁刺激(rTMS)第一天 H1 型线圈对空间识别记忆有短暂的短期影响(但在后续治疗日没有)。另一方面,H2 型线圈治疗使空间工作记忆得到短暂改善。最后,问卷调查显示除了接受 H1 型线圈治疗的受试者报告有“分离感”外,没有显著的情绪或心境改变。

结论

本研究为两种 H 型线圈设计(H1/H2)的可行性和安全性提供了更多证据。

意义

H 型线圈提供了一种安全的新工具,在与深部脑区功能障碍相关的精神和神经疾病的研究及临床应用方面具有潜力。