Department of General Intensive Care, St James Wing, St George's Hospital, Tooting, London, UK.
Minerva Anestesiol. 2010 Dec;76(12):1010-7. Epub 2010 Jul 16.
Pulse pressure (PP) analysis from a radial arterial line is available with the LiDCO plus monitor (LiDCO, Cambridge, UK) and FloTrac/Vigileo (Edwards Lifesciences, Irvine, CA, USA). The aim of this study was to investigate the agreement of continuous PP analysis against intermittent thermodilution (ITD) using the pulmonary artery catheter (PAC).
This was a six-hour study in 29 patients monitored with a PAC. All measurements were referenced against CO measured from the average of four ITD curves from the PAC. The LiDCO plus was calibrated with a lithium dilution (PulseCOLi) and with ITD (PulseCOITD) at baseline. Measurements from Vigileo software 1.03 (APCO), LiDCO plus (PulseCOLi and PulseCOITD), CCO and ITD were taken every hour for the next six hours. The bias and precision between the two devices were calculated as well as the percentage error (PE) of agreement between the tested device and the reference. The coefficient of variation (CV) of the tested device was then derived.
The average bias, PE and coefficient of variation for CCO vs. ITD of the tested device were 0.3 L/min, 28% and 13%, respectively; for APCO vs. ITD the calculations were -1.1 L/min, 55% and 27%; for PulseCOLi Cardiac output Blood pressure Thermodilution. ITD they were 0.5 L/min, 40% and 19%; and for PulseCOITD vs. ITD they were 0.2 L/min, -31% and 15%.
APCO (Vigileo software 1.03) and PulseCOLi showed a moderate agreement with the PAC. When PulseCO was calibrated with ITD (PulseCOITD) it showed excellent agreement, demonstrating that PulseCO performs well against ITD when the calibration process is optimally performed.
来自桡动脉导管的脉搏压 (PP) 分析可通过 LiDCO plus 监测仪(LiDCO,英国剑桥)和 FloTrac/Vigileo(爱德华生命科学公司,美国加利福尼亚州欧文)获得。本研究旨在探讨连续脉搏压分析与肺动脉导管 (PAC) 间歇性热稀释法 (ITD) 的一致性。
这是一项在 29 例接受 PAC 监测的患者中进行的六小时研究。所有测量均参考来自 PAC 四条 ITD 曲线的平均 CO。LiDCO plus 通过锂稀释(PulseCOLi)和 ITD(PulseCOITD)在基线时进行校准。Vigileo 软件 1.03(APCO)、LiDCO plus(PulseCOLi 和 PulseCOITD)、CCO 和 ITD 的测量值每小时测量一次,持续六小时。计算两种设备之间的偏差和精度以及测试设备与参考设备之间的百分比误差 (PE)。然后得出测试设备的变异系数 (CV)。
CCO 与测试设备的 ITD 相比,APCO 的平均偏差、PE 和 CV 分别为 0.3 L/min、28%和 13%;APCO 与 ITD 相比,计算值分别为-1.1 L/min、55%和 27%;PulseCOLi 心输出量血压热稀释法。与 ITD 相比,ITD 的计算值分别为 0.5 L/min、40%和 19%;与 ITD 相比,PulseCOITD 的计算值分别为 0.2 L/min、-31%和 15%。
APCO(Vigileo 软件 1.03)和 PulseCOLi 与 PAC 具有中等一致性。当 PulseCO 通过 ITD(PulseCOITD)进行校准时,它表现出极好的一致性,表明当校准过程得到优化时,PulseCO 与 ITD 配合良好。