The purpose of this study was to assess the reproducibility of resting blood flow, resting oxygen consumption, and mitochondrial capacity in skeletal muscle using near-infrared spectroscopy (NIRS). We also determined the influence of 2 exercise modalities (ergometer and rubber exercise bands) on the NIRS measurements. Fifteen young, healthy participants (5 female, 10 male) were tested on 2 nonconsecutive occasions within an 8-day period. The NIRS device was placed on the medial gastrocnemius. Venous and arterial occlusions were performed to obtain blood flow and oxygen consumption. A series of repeated arterial occlusions was used to measure the recovery kinetics of muscle oxygen consumption after ∼7-10 s of voluntary plantar flexion exercise. Resting blood flow had mean coefficients of variation (CV) of 42% and 38% for bands and ergometer, respectively, and resting metabolism had mean CVs of 17% and 12% for bands and ergometer, respectively. The recovery time constant of oxygen consumption (day 1 bands and ergometer: 23.2 ± 3.7 s, 27.6 ± 6.5 s, respectively; day 2 bands and ergometer: 25.5 ± 5.4 s, 25.0 ± 4.9 s, respectively) had mean CVs of 10% and 11% for bands and ergometer, respectively. We conclude that measurements of oxygen consumption and mitochondrial capacity using NIRS can be obtained with good reproducibility.
本研究的目的是使用近红外光谱(NIRS)评估骨骼肌静息血流量、静息氧消耗和线粒体容量的可重复性。我们还确定了两种运动方式(测力计和弹力带运动)对NIRS测量的影响。15名年轻健康的参与者(5名女性,10名男性)在8天内分两次非连续进行测试。NIRS设备放置在内侧腓肠肌上。通过静脉和动脉阻塞来获取血流量和氧消耗。使用一系列重复的动脉阻塞来测量在约7 - 10秒的主动跖屈运动后肌肉氧消耗的恢复动力学。静息血流量方面,弹力带组和测力计组的平均变异系数(CV)分别为42%和38%,静息代谢方面,弹力带组和测力计组的平均CV分别为17%和12%。氧消耗的恢复时间常数(第1天弹力带组和测力计组分别为:23.2±3.7秒、27.6±6.5秒;第2天弹力带组和测力计组分别为:25.5±5.4秒、25.0±4.9秒),弹力带组和测力计组的平均CV分别为10%和11%。我们得出结论,使用NIRS测量氧消耗和线粒体容量具有良好的可重复性。