Baek Jiwon, Doh Sang Hee, Chung Sung Kun
*Department of Ophthalmology and Visual Science, Seoul St Mary's Hospital, College of Medicine, The Catholic University of Korea; and †Department of Ophthalmology and Visual Science, St Paul's Hospital, College of Medicine, The Catholic University of Korea.
Cornea. 2015 Oct;34(10):1209-13. doi: 10.1097/ICO.0000000000000575.
To study the repeatability and reproducibility of tear meniscus height (TMH) measurements obtained with a keratograph and Fourier domain optical coherence tomography (FD-OCT) and to assess their agreement in patients with dry eye.
Sixty-four eyes with dry eye were analyzed by the Schirmer test, tear breakup time test, and fluorescein corneal staining. The TMH was measured 3 times using both devices by 2 different examiners. The repeatability and reproducibility of measurements were assessed by within-subject standard deviation (Sw), repeatability (2.77 Sw), coefficient of variation (CoV), and intraclass correlation coefficient.
The TMH measured with the keratograph and FD-OCT was 0.232 ± 0.074 mm and 0.308 ± 0.129 mm, respectively (P < 0.01). A close correlation was found between the TMH measured with the keratograph and FD-OCT (r = 0.343). There was a negative correlation between the mean TMH and differences in the TMH measured with the keratograph and FD-OCT (r = 0.359). Both measurements correlated with the Schirmer test score, tear breakup time, and corneal staining score with P < 0.01. Intraexaminer CoV, 2.77 Sw, and intraclass correlation coefficient of the TMH were <6.5%, <0.059 mm, and >0.986, respectively, and interexaminer CoV and 2.77 Sw were 5.58% and 0.039 mm, respectively.
Although the TMH measured with the keratograph tended to be lower than that measured with FD-OCT, the TMH measured with the keratograph closely correlated with the TMH measured with FD-OCT and conventional tests and had good repeatability and reliability.
研究使用角膜地形图仪和傅里叶域光学相干断层扫描(FD-OCT)测量泪液半月板高度(TMH)的可重复性和再现性,并评估其在干眼患者中的一致性。
对64只干眼患眼进行泪液分泌试验、泪膜破裂时间试验和荧光素角膜染色分析。由2名不同的检查者使用这两种设备对TMH进行3次测量。通过受试者内标准差(Sw)、重复性(2.77Sw)、变异系数(CoV)和组内相关系数评估测量的可重复性和再现性。
角膜地形图仪和FD-OCT测量的TMH分别为0.232±0.074mm和0.308±0.129mm(P<0.01)。角膜地形图仪和FD-OCT测量的TMH之间存在密切相关性(r=0.343)。平均TMH与角膜地形图仪和FD-OCT测量的TMH差异之间存在负相关(r=0.359)。两种测量均与泪液分泌试验评分、泪膜破裂时间和角膜染色评分相关,P<0.01。TMH的检查者内CoV、2.77Sw和组内相关系数分别<6.5%、<0.059mm和>0.986,检查者间CoV和2.77Sw分别为5.58%和0.039mm。
虽然角膜地形图仪测量的TMH往往低于FD-OCT测量的TMH,但角膜地形图仪测量的TMH与FD-OCT测量的TMH以及传统检查密切相关,具有良好的可重复性和可靠性。